Model Number SIGTRSB60AMT |
Device Problems
Misfire (2532); Separation Failure (2547)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/04/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the device was partially fired.It was reported that after firing, the reinforcement material did not release from the device as expected.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The evaluation detected an unreported condition: the device had damage to the cutting edge of the knife blade.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.D10 concomitant product: egiaustnd, egiaustnd endogia ultra univ std stap (lot#unk) medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during the functional end to end anastomosis on a laparoscopic colectomy, at the time of doing the side to side anastomosis, the anchoring suture on the tip did not come off after firing, so gentle tension was applied and the sheet was detached.There was no insufficient squeezing, and it was squeezed, as well as checked, until the end.The handle was still used afterwards, but there were no problem, so it was believed to be caused by the cartridge.There was no patient injury.Medtronic's initial evaluation of the incident device found that the reload was partially fired and the interlock was engaged.
|
|
Search Alerts/Recalls
|
|
|