The reported events were lead dislodgement, oversensing noise, inappropriate shock, and helix mechanism issue.As received, a complete lead was returned in one piece for analysis.The reported events of oversensing noise and inappropriate shock were not confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual inspection of the lead body did not find any anomalies.X-ray examination found the inner coil over-torqued at the connector region consistent with procedural damage.The reported event of helix mechanism issue was confirmed.Visual inspection found the helix to be retracted and clogged with blood.After cutting the lead and cleaning the distal portion of the lead, the helix could be extended and retracted by applying torque directly to the inner coil.The measured full helix extension length was within specification.The cause of helix mechanism issue was isolated to the helix clogged with blood, blood on the inner coil, and over-torque of the inner coil.
|