• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH EARLYVUE VS30 VITAL SIGNS MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH EARLYVUE VS30 VITAL SIGNS MONITOR Back to Search Results
Model Number 863359
Device Problem No Audible Prompt/Feedback (2282)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/28/2023
Event Type  malfunction  
Manufacturer Narrative
A follow up report will be submitted once the investigation is complete.
 
Event Description
It was reported that a speaker malfunction error was displayed.It is unknown if there was still sound coming from the device.It is unknown if the device was in use at time of event and no patient harm was reported.
 
Manufacturer Narrative
A philips remote service engineer (rse) interviewed the customer onsite to evaluate the devices in question.The customer reported that despite having already replaced the speaker, the technical alarm message remains.The rse troubleshot with the customer and confirmed the unit the unit gave a speaker inop message as soon as the monitor was turned on.The philips field service engineer (fse) confirmed there was no issues with the speaker and it was fully functional.The unit was tested and the issue was confirmed to be the mother board.The customer requested a quote to order a new mother board.The motherboard was replaced and returned to functional use with no further issues identified.The device remains at the customer site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EARLYVUE VS30 VITAL SIGNS MONITOR
Type of Device
EARLYVUE VS30 VITAL SIGNS MONITOR
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key16817079
MDR Text Key314028942
Report Number1218950-2023-00261
Device Sequence Number1
Product Code DSJ
UDI-Device Identifier00884838091412
UDI-Public00884838091412
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K190624
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number863359
Device Catalogue Number863359
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/28/2023
Initial Date FDA Received04/26/2023
Supplement Dates Manufacturer Received05/24/2023
Supplement Dates FDA Received06/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-