• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INSIGHTEC LTD. EXABLATE 4000; MR GUIDED FOCUSED ULTRASOUND SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSIGHTEC LTD. EXABLATE 4000; MR GUIDED FOCUSED ULTRASOUND SYSTEM Back to Search Results
Model Number 4000
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Burn(s) (1757)
Event Date 03/14/2023
Event Type  Injury  
Manufacturer Narrative
Known risk of the device.No new risk has been recognized.
 
Event Description
Skin burns at stereotactic frame pin sites following essential tremor treatment.
 
Manufacturer Narrative
Known risk of the device.No new risk has been recognized.No device malfunction.Skin burns healed completely and therefore event was transient.Since burns only appeared two days after treatment, and were only present at two out of four contact points with the skin, it is possible that the adverse event is not device related.
 
Event Description
Skin burns at stereotactic frame pin sites on the forhead were seen two days after essential tremor treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXABLATE 4000
Type of Device
MR GUIDED FOCUSED ULTRASOUND SYSTEM
Manufacturer (Section D)
INSIGHTEC LTD.
5 nahum heth street
tirat carmel, 39120
IS  39120
Manufacturer (Section G)
INSIGHTEC LTD.
5 nachum heth
tirat carmel, 39120
IS   39120
Manufacturer Contact
may zimmerman
5 nachum heth
tirat carmel, 39120
IS   39120
MDR Report Key16822056
MDR Text Key314074812
Report Number9615058-2023-00013
Device Sequence Number1
Product Code POH
UDI-Device Identifier07290015461021
UDI-Public010729001546102112109022140101
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model Number4000
Device Catalogue NumberSYS940350
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/27/2023
Initial Date FDA Received04/27/2023
Supplement Dates Manufacturer Received03/27/2023
Supplement Dates FDA Received06/15/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Disability;
Patient SexMale
-
-