Model Number ACT200 |
Device Problems
Device Alarm System (1012); Contamination (1120); Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/13/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that after the use of an act plus instrument, it was reported that the instrument had a measurement issue, and it was giving a wrong measurement.There was no adverse patient effect associated with this event.Medtronic received additional information that the customer confirmed that the failed quality control (qc) measurement, was the liquid control and that the actuator assembly was found dirty with blood.The customer complained of measurements only in one cell.The following errors were found in the statistic log file: 010, 012,015.
|
|
Manufacturer Narrative
|
Device evaluation summary: the reported measurement issue was verified during service.Service technician observed errors 010, 012,015 (routine errors) in the statistic error log.Actuator assembly was dirty with blood, and preventive maintenance was overdue.The issue was resolved by cleaning the instrument, and the reagent drive slider block, cleared statistic log file (no serious errors), adjusted the lift wire to within specification, performed acttrac cartridge test and result was ok.Performed operation with clottrac hr normal coagulation controls and deionized water (test liquid), results o.K.Preventive maintenance was performed as per specification.Conclusions: after investigation at medtronic, the complaint was confirmed for the act plus instruments measurement issue, and giving the wrong measurement.No patient/clinical safety issues reported.Trends for issues with this product are reviewed at quarterly quality meetings.The act plus software constantly monitors for software and/or hardware faults.When one is found, an error is displayed and typically addressed by the operator.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|