• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DENALI FEMORAL SYSTEM; VENA CAVA FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DENALI FEMORAL SYSTEM; VENA CAVA FILTER Back to Search Results
Catalog Number DL950F
Device Problems Break (1069); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/13/2023
Event Type  malfunction  
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the denali filter products that are cleared in the us.The pro code and 510k number for the denali filter products is identified in d2 and g4.H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiry date: 06/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during a inferior vena cava filter placement procedure, it was allegedly found that the guidewire in the push rod was broken which caused the filter unable to be deployed.The procedure was completed using another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the denali filter products that are cleared in the us.The pro code and 510k number for the denali filter products is identified in d2 and g4.H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported break issue could not be determined based upon the provided information labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 06/2025), g3 h11: h6 (device) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during a inferior vena cava filter placement procedure, it was allegedly found that the guidewire in the push rod was broken which caused the filter unable to be deployed.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DENALI FEMORAL SYSTEM
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key16825432
MDR Text Key314115127
Report Number2020394-2023-00297
Device Sequence Number1
Product Code DTK
UDI-Device Identifier00801741040825
UDI-Public(01)00801741040825
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K130366
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberDL950F
Device Lot NumberGFGT3156
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/03/2023
Initial Date FDA Received04/27/2023
Supplement Dates Manufacturer Received06/13/2023
Supplement Dates FDA Received06/15/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-