BAXTER HEALTHCARE CORPORATION MINICAP; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number 5C4466P |
Device Problems
Disconnection (1171); Fitting Problem (2183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The event occurred on three (3) unspecified dates of (b)(6) 2023.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that three (3) minicaps were ¿not securing properly¿ to the transfer set; the minicaps were "falling off no matter how tight they attempted to secure the minicap¿.This occurred during use of the devices for peritoneal dialysis (pd) therapy.One (1) of the caps fell off and was not noticed until a "couple of hours".The other caps did not secure properly during their initial attempt.The transfer set remained in use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The devices were discarded; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
|