Model Number CDT2212
Device Problems Migration or Expulsion of Device (1395); Migration (4003)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2023
Event Type  malfunction  
Event Description
On (b)(6) 2023, due to descending colon obstruction, the patient was implanted in the gastroscopy room of (b)(6) with niti-s enteral colonic uncovered stent, cdt2212.After the stent was released, feces gushed out and blocked the endoscopic field of vision.After rinsing, the stent was found to have shifted inward without support at the focal site, and then the second niti-s enteral colonic uncovered stent was implanted.
 
Manufacturer Narrative
It was reported that after stent was released, feces gushed out and the stent was found to have shifted from the focal site.It was confirmed from the device history record that device had been manufactured with no significant issue and passed all the inspections successfully.Migration can occur in any company's device as well as ours.It is affected by patient's lesion status, peristalsis of organs, and drug use in general.It is difficult to identify the exact root cause because the device was not returned, and it is difficult to reconstruct the situation at the time of procedure.However, based on the description "after the stent was released, feces gushed out and blocked the endoscopic field of vision.After rinsing, the stent was found to have shifted inward without support at the focal site", it is assumed the stent was migrated due to progression of the patient's disease and patient's condition, feces during the procedure and other factors complexly.Through the user manual by taewoong, it is stated that "potential complications associated with the use of niti-s & comvi stent may include, but are not limited to: stent migration".This suspected device is not registered in the us but we will continuously monitor the same or similar customer complaints through accurate analyses.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NITI-S ENTERAL COLONIC UNCOVERED STENT
Type of Device
COLONIC STENT
Manufacturer (Section D)
TAEWOONG MEDICAL CO.,LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
KS  10022
Manufacturer (Section G)
TAEWOONG MEDICAL CO.,LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
KS   10022
Manufacturer Contact
lee
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
MDR Report Key16828668
MDR Text Key314297078
Report Number3003902943-2023-00012
Device Sequence Number1
Product Code MQR
Combination Product (y/n)Y
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDT2212
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/13/2023
Initial Date FDA Received04/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-