C.R. BARD, INC. (BASD) -3006260740 M.R.I. LOW-PROFILE IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 7F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 0603870CE |
Device Problems
Fracture (1260); Material Separation (1562); Dent in Material (2526)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/03/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that approximately one year and two months post port catheter placement, during removal a large part of the catheter allegedly broke and remained in the vein.It was further reported that the catheter was removed.Reportedly, the patient experienced pain in the right clavicle but there is no intervention or medication prescribed for the treatment for pain.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the m.R.I.Low-profile implantable port, groshong single-lumen, 7f products that are cleared in the us.The pro code and 510k number for the m.R.I.Low-profile implantable port, groshong single-lumen, 7f products is identified in d2 and g4.H10: as the lot number for the device was provided, a review of the device history records will be performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiry date: 01/2025).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the m.R.I.Low-profile implantable port, groshong single-lumen, 7f products that are cleared in the us.The pro code and 510k number for the m.R.I.Low-profile implantable port, groshong single-lumen, 7f products is identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one bard implantable port attached to a catheter was received.Visual, microscopic and functional evaluations were performed.A complete circumferential break was noted on the distal end of the attached catheter that appeared elliptical in shape.The edges of the break was noted to be uneven and the surface appeared to be granular.The distal catheter segment was not returned.Therefore, the investigation is confirmed for the reported fracture, material separation and the identified dent issues.Although a definitive root cause could not be determined, pinch off could have potentially caused or contributed to the reported event.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 01/2025), g3, h6 (device) h11: b3, b5, h6 (method, result, conclusion) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that sometimes post a port placement, during removal a large part of the catheter allegedly broke and remained in the vein.It was further reported that the catheter was removed.Reportedly, the patient experienced pain in the right clavicle but there is no intervention or medication prescribed for the treatment for pain.There was no reported patient injury.
|
|