SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number CA32B0/400/004JP |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that an air leak alarm was detected during the pre-use check.It was reported that there no additional information is available for this complaint.
|
|
Manufacturer Narrative
|
B3: month and year of event have been provided, day is unknown.D4: expiration date and h4: manufacture date are unknown, no information is available based on reported lot number.G5: no 510k as device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
One used device (pre-use check) was received.Visual inspection noted a black marking by the customer and a tear was found in the center of the corrugate tube.A leak was detected during functional testing confirming the complaint.A root cause could not be determined however the damage could have occurred during transportation and/or handling of the product.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.This issue will be monitored, and further actions taken accordingly.
|
|