SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number CA17K0/000/004JP |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that during the pre-use check, leakage of air from the product was observed.The customer checked it again by submerging it in water, and air bubbles were found at the connection part (where the breathing filter and the breathing circuit were connected).No patient involvement.
|
|
Manufacturer Narrative
|
B3: date of event and d4: udi section are unknown, no information has been provided to date.D4: expiration date h4: manufacture date are not available based on the reported lot number.G5: 510k is blank, device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
|
|
Manufacturer Narrative
|
Updated one device was received for investigation.A leak test was conducted, but no leakage was found.The complaint is not confirmed.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
|
|