• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Catalog Number CA17K0/000/004JP
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that during the pre-use check, leakage of air from the product was observed.The customer checked it again by submerging it in water, and air bubbles were found at the connection part (where the breathing filter and the breathing circuit were connected).No patient involvement.
 
Manufacturer Narrative
B3: date of event and d4: udi section are unknown, no information has been provided to date.D4: expiration date h4: manufacture date are not available based on the reported lot number.G5: 510k is blank, device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
 
Manufacturer Narrative
Updated one device was received for investigation.A leak test was conducted, but no leakage was found.The complaint is not confirmed.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16835216
MDR Text Key314223337
Report Number3012307300-2023-04747
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCA17K0/000/004JP
Device Lot Number239207
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/13/2023
Initial Date FDA Received04/28/2023
Supplement Dates Manufacturer Received06/29/2023
Supplement Dates FDA Received06/30/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-