ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM (ACTIVE SENTRY); UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065753057
Device Problems Inability to Irrigate (1337); No Pressure (2994)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/30/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that during a cataract procedure a system message was displayed and there was loss in pressure in the chamber.The irrigation was stopped in irrigation/aspiration (ia) mode and resumed after clearing the system message.Procedure was completed by replacing the surgical procedure pack.There was no patient harm.
 
Manufacturer Narrative
Service history was reviewed for the system.No service record relevant to the complaint reported event was found.However, the system was last serviced prior to the reported event per service record (sr) opened.The system found to meet all cosmetic and performance standards.There was no samples returned for testing on this investigation.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.The root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM (ACTIVE SENTRY)
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16841284
MDR Text Key314264253
Report Number2028159-2023-00620
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161794
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065753057
Device Lot Number13MAL6
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/04/2023
Initial Date FDA Received04/29/2023
Supplement Dates Manufacturer Received06/29/2023
Supplement Dates FDA Received07/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-