• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-H180J
Device Problems Contamination /Decontamination Problem (2895); Failure to Clean Adequately (4048)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/31/2023
Event Type  malfunction  
Manufacturer Narrative
The customer suspected device was returned for investigation.The olympus service center confirmed the reported event during inspection and testing.Upon evaluation of the returned device the following defects were found, no air water flow detected due to clogged nozzle with foreign object, after clogging was removed, air water flow was normal, distal end cover insulation value equal to zero, distal end cover dented, angle rubber glue cracked, objective lens chipped, objective lens glue had hole, light guide lens glue cracked, angulation low, control knob paly, right/left/up/down knob heavy when engage and label illegible peeling.The faulty parts will be replaced, and the device will be returned to the user facility.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus that during reprocessing of the evis exera ii gastrointestinal videoscope, the scope had irrigation issue.Upon inspection and testing of the returned device, foreign material was found clogged in the scope air/water nozzle due to insufficient cleaning.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.There was no patient involvement reported with this event.
 
Manufacturer Narrative
Updated: h6, h10.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the observed foreign matter clogging the nozzle could not be identified and a definitive root cause of the issue could not be determined, as there was no device deformation the could have contributed to the retention of foreign material and no additional event or device reprocessing information was reported.The event can be detected/prevented by following the instructions for use sections below: evis exera ii gif/cf/pcf type 180 series operation manual chapter 3 preparation and inspection.Evis exera ii gif/cf/pcf type 180 series reprocessing manual chapter 5 reprocessing the endoscope (and related reprocessing accessories) describes preventive measures).Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16850273
MDR Text Key314560488
Report Number9610595-2023-06954
Device Sequence Number1
Product Code FDS
UDI-Device Identifier04953170275456
UDI-Public04953170275456
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051645
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-H180J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/03/2023
Initial Date FDA Received05/01/2023
Supplement Dates Manufacturer Received07/03/2023
Supplement Dates FDA Received07/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/12/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-