Model Number L301 |
Device Problems
Signal Artifact/Noise (1036); Failure to Capture (1081); High impedance (1291); Over-Sensing (1438); Incorrect, Inadequate or Imprecise Result or Readings (1535); Low impedance (2285); Inappropriate or Unexpected Reset (2959); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/03/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that this pacemaker system exhibited noise with oversensing on the right atrial (ra) lead and noise on the right ventricular (rv) lead with possible crosstalk oversensing.In addition, one atrial tachy response (atr) episode showed a single beat of rv loss of capture (loc).Technical services (ts) reviewed troubleshooting options.This pacemaker system remains in service.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that this pacemaker system exhibited noise with oversensing on the right atrial (ra) lead and noise on the right ventricular (rv) lead with possible crosstalk oversensing.In addition, one atrial tachy response (atr) episode showed a single beat of rv loss of capture (loc).Technical services (ts) reviewed troubleshooting options.This pacemaker system remains in service.No adverse patient effects were reported.Additional information received reported that this pacemaker had entered safety mode and recorded codes 0xa 0xa 0xa, indicative of an oscillator mismatch.There was a continuation of noise observed on both the right atrial (ra) and right ventricular (rv) leads.Review of ra impedance trends showed impedance fluctuations between 500 and 1000 ohms, and rv impedance trends indicated a decrease in pace impedance measurements from 600 ohms to as low as 481 ohms.Additionally, rv threshold measurements showed occasionally spikes from 0.9v to 3v.The pacemaker was explanted and replaced, and the rv lead was surgically abandoned and replaced during the same procedure.The ra lead remains in service.No additional adverse patient effects were reported.This pacemaker is expected to be returned for analysis.
|
|
Search Alerts/Recalls
|