Model Number CDT1808
Device Problem Migration (4003)
Patient Problems Pain (1994); Perforation (2001)
Event Date 04/13/2023
Event Type  Injury  
Manufacturer Narrative
It was reported that 1 hour after stent placement, patient complained of pain and stent was found migrated and perforated in gastro intestinal tract.It was confirmed from the device history record that device had been manufactured with no significant issue and passed all the inspections successfully.Migration can occur in any company's device as well as ours.It is affected by patient's lesion status, peristalsis of organs, and drug use in general.In addition, perforation can occur due to other company's device as well as ours.It is affected by patient's lesion status, peristalsis of organs, and drug use in general.In addition, however, it is hard to identify the exact root cause since it is hard to reconstruct the situation at the time of procedure.It is difficult to identify the exact root cause because the device was not returned, and it is difficult to reconstruct the situation at the time of procedure.However, based on the description "stent placement would not be stable because the tumor was not coming into git", it is assumed the stent was migrated and perforated at the gastro intestinal tract due to progression of the patient's disease and patient's condition, peristalsis of organs and other factors complexly.Through the user manual by taewoong, it is stated that "potential complications associated with the use of niti-s & comvi stent may include, but are not limited to: stent migration, perforation".This suspected device is not registered in the us but we will continuously monitor the same or similar customer complaints through accurate analyses.
 
Event Description
This patient is not given chemotherapy and the stent was not all circumstance stenosis.The patient complained of pain after approx.1 hour of stent placement, and then stent migration and perforation in git.It is unknown if the perforation was caused by the stent.The perforation may be at the tumor and the stent was not fractured in this event.The physician understood that the stent placement would not be stable because the tumor was not coming into git.There were no patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NITI-S ENTERAL COLONIC UNCOVERED STENT
Type of Device
COLONIC STENT
Manufacturer (Section D)
TAEWOONG MEDICAL CO.,LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, 10022
KS  10022
Manufacturer (Section G)
TAEWOONG MEDICAL CO.,LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, 10022
KS   10022
Manufacturer Contact
byeoungcheol lee
14, gojeong-ro
wolgot-myeon
gimpo-si, 10022
KS   10022
MDR Report Key16850983
MDR Text Key314391503
Report Number3003902943-2023-00015
Device Sequence Number1
Product Code MQR
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/07/2024
Device Model NumberCDT1808
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/14/2023
Initial Date FDA Received05/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/08/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-