Model Number 1365-22-000 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Fluid Discharge (2686)
|
Event Date 07/07/2021 |
Event Type
Injury
|
Event Description
|
Medical records : on follow-up visit: 7-jul-2021: patient underwent a revision on (b)(6) 2021 of all implants.During removal of the stem, the calcar was fractured requiring two cables to fixate the bone.Postoperatively, excessive drainage from the wound occurred requiring a wound vac to be placed.Doi: (b)(6) 2013, dor: (b)(6) 2021, affected side: right hip, this pc is linked to (b)(4) ( main pc ).
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).E3 initial reporter occupation: lawyer.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.Through follow up it is now understood there is no allegation associated against a product malfunction.Review of the x-ray evidence found nothing indicative of a device nonconformance or a relation of the product with the reported adverse event.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Search Alerts/Recalls
|