Model Number DTBC2D1
Device Problem Misconnection (1399)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/27/2023
Event Type  malfunction  
Event Description
It was reported that the right ventricular (rv) lead exhibited suddenly increased out of range high impedance values on the superior vena cava (svc) and rv defib coils.A connection issue was suspected with the cardiac resynchronization therapy defibrillator (crt-d) device.A revision was performed.It was noted that there was liquid in the connector.After cleaning and testing the impedances were back to normal.The device and leads are still in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAVA¿ CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key16860330
MDR Text Key314572096
Report Number9614453-2023-01538
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeRS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/14/2023
Device Model NumberDTBC2D1
Device Catalogue NumberDTBC2D1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/28/2023
Initial Date FDA Received05/03/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/27/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
694765 LEAD.
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age61 YR
Patient SexMale
-
-