PHILIPS MEDICAL SYSTEMS PATIENT INFORMATION CENTER IX Back to Search Results
Model Number 866389
Device Problem Application Network Problem (2879)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/09/2023
Event Type  malfunction  
Event Description
Philips received a complaint on the piic ix indicating that disconnecting from server and unable to print.The device was in use.There were no adverse events involving patients.
 
Manufacturer Narrative
Visual inspection found multiple surveillance in local mode.The complaint was escalated for technical investigation and the results indicate multiple surveillance are in local mode and including multiple switches and core a showing off line.Based on the information available and the testing conducted, the cause of the reported problem was core was dropping traffic.The reported problem was confirmed.The device was operational after repairs were completed.The investigation concludes that the customer is taking responsibility to correct/repair the device.The customer has been notified to contact philips if this does not address their device issue at which point a new service order will be created.No further information will be obtained as this concluded philips remote support to the customer for this event.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATIENT INFORMATION CENTER IX
Type of Device
PATIENT INFORMATION CENTER IX
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16860638
MDR Text Key314462657
Report Number1218950-2023-00287
Device Sequence Number1
Product Code MHX
UDI-Device Identifier00884838093041
UDI-Public00884838093041
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153702
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866389
Device Catalogue Number866389
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/09/2023
Initial Date FDA Received05/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/05/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-