No device was returned to for evaluation and no photographs were provided for review.As a result, the issue was unable to be confirmed with the information provided.No further investigation could be completed.Based on the information obtained the root cause of the reported event is unknown but may have been the result of patient preparation or other intraoperative environmental factors.Labeling review: "precaution: while the nvm5 system is designed to assist in the electromyographic location of spinal nerves in proximity to the surgical site, it is not intended to take the place of thorough knowledge of spinal anatomy and appropriate surgical technique, nor should the information provided by the system be construed as definitive indicators of nerve location.Such factors as the distance from the nerve, the position and placement of electrodes, individual muscle and/or nerve responses, the proximity and strength of sources of electrical interference, and other patient and environmental factors, may influence the operation.If, in the judgment of the clinician, this resistance is sufficient to preclude proper placement of instruments, the procedure should be suspended." "a thorough cleaning and preparation of the dermal surface prior to placement of recording electrodes is required for proper adherence and sensitivity of the electrodes." ".Using the provided electrode placement instructions.Failure to follow these instructions may result in the display of inadequate information necessary for data interpretation.".
|