• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS PATIENT INFORMATION CENTER IX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS PATIENT INFORMATION CENTER IX Back to Search Results
Model Number 866389
Device Problem Application Network Problem (2879)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/10/2023
Event Type  malfunction  
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Event Description
The customer reported that there is a system wide shut down every month where all hosts going into local for 3-5 minutes.The device was in use.There was no report of patient or user harm.
 
Manufacturer Narrative
Logs have been pulled from multiple locations.This case was escalated to the national support specialist (nss) team.All centrals and servers lose connection to the server at the same time.Multiple fse¿s have been onsite to collaborate with the customer on this hybrid network; two hospitals have a philips supplied network and servers are customer supplied network.It was confirmed that this was not a planned or scheduled outage.The customers¿ servers were scheduled to back up once a month.Due to being a server issue, the customer was to fix the issue.The fse¿s gave them what they thought was the cause of the issue and they would have to take action to change on their end.The device was confirmed to be operating per specifications and no failure was identified.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATIENT INFORMATION CENTER IX
Type of Device
PATIENT INFORMATION CENTER IX
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16868310
MDR Text Key314552016
Report Number1218950-2023-00290
Device Sequence Number1
Product Code MHX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153702
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866389
Device Catalogue Number866389
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/10/2023
Initial Date FDA Received05/04/2023
Supplement Dates Manufacturer Received04/10/2023
Supplement Dates FDA Received05/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/18/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-