Model Number CLV-190
Device Problems Power Problem (3010); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/10/2023
Event Type  malfunction  
Event Description
The customer reported to olympus, that the evis exera iii xenon light source's fan motor was worn out.It was added, that the device was making noises and that the electrical board lights were all blinking.There was no report of patient harm.
 
Manufacturer Narrative
The device was returned to olympus for evaluation of the reported issue.The allegation regarding the fan making noise was not confirmed.However, the blinking lights were confirmed, due to a faulty socket slider switch.Other findings include a cracked front panel mouth, a rusted chassis, and a noisy air pump.Additionally, the device's non-olympus lamp life meter was reading at 499 hours and 32 minutes.The light output measured out of range.And a replacement to the olympus xenon lamp was recommended for optimal performance.The investigation is ongoing.And a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation, and to correct information provided on the initial report.Based on the results of the investigation, it was determined that phenomenon, ¿all lights/lamps blink" was reproduced, and it was due to a slider switch failure in the socket.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 years since the subject device was manufactured.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16872714
MDR Text Key314641541
Report Number3002808148-2023-04559
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/10/2023
Initial Date FDA Received05/04/2023
Supplement Dates Manufacturer Received05/31/2023
Supplement Dates FDA Received06/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/27/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-