TIDI PRODUCTS LLC POSEY BED 8070; PATIENT BED WITH CANOPY/RESTRAINTS Back to Search Results
Model Number 8070
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
H3 evaluation of the canopy found a zipper that does not engage correctly in a patient access area.A zipper that does not engage correctly may allow the zipper to be initially closed.A patient can later open the zipper by applying pressure to the zipper at the location of the misaligned zipper elements (teeth).In this case, there was no impact or consequence to the patient and the canopy was returned for servicing when the zipper issue was identified.Although it cannot be confirmed, it is possible that routine wear-and-tear from repeated use contributed to the zipper damage.Posey beds are multi-use, serviceable items.As such, it is anticipated that the units may occasionally require repair, and as part of standard care the beds should be inspected prior to use.If damage is noted during these routine bed inspections, the unit should not be put into use with a patient and should be returned to posey for servicing.Instructions for use (ifu) were reviewed and found to provide adequate instructions and warnings for safe and effective use of the device.Service issues are trended and reviewed by management on a monthly basis.As part of this monthly review, trends and excursions above control limits will be assessed, documented and acted upon as warranted.No corrective or preventative actions are necessary at this time manufacturer reference file (b)(4).
 
Event Description
(b)(6) reported via email there are canopies in need of repair.Customer owns product,return repaired.Sn (b)(6) broken zipper.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POSEY BED 8070
Type of Device
PATIENT BED WITH CANOPY/RESTRAINTS
Manufacturer (Section D)
TIDI PRODUCTS LLC
570 enterprise drive
neenah WI 54956
Manufacturer Contact
chris rahn
570 enterprise drive
neenah, WI 54956
9207514300
MDR Report Key16872857
MDR Text Key314831618
Report Number2182318-2023-00045
Device Sequence Number1
Product Code OYS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113357
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8070
Device Catalogue Number8070
Device Lot Number2019082834
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2023
Initial Date Manufacturer Received 03/02/2023
Initial Date FDA Received05/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/29/2019
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-