• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP INC. PRESTIGE ATRA GRASPER DBL-ACT 5MM; HANDHELD PRODUCTS & LIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP INC. PRESTIGE ATRA GRASPER DBL-ACT 5MM; HANDHELD PRODUCTS & LIGATION Back to Search Results
Model Number 8360-10
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event Description
It was reported to aesculap inc.That a prestige atra grasper dbl-act 5mm (product # 8360-10) was used during a procedure on an unspecified date.According to the complainant the jaw of the device fell apart into multiple pieces.A delay to surgery was reported.No additional information has been provided although it has been requested.The adverse event is filed under aic reference (b)(4).
 
Manufacturer Narrative
Distal weld: visual inspection summary: 8360-10 article has no jaws within the jaw assembly.Components appear broken and the jaws have either broken or disassembled from the device.The device was visually inspected and the complaint of "broke during surgery" can be confirmed as the jaws disassembled from the device.No further testing will be performed as the device does not function as intended in its current state.The investigation into the cause of the reported problem was able to confirm the failure mode of distal jaw separation.In addition to a supplier corrective action request (scar) being initiated, a corrective action/preventive action (capa) was opened by aesculap inc.For further evaluation of the design transfer of this device.
 
Event Description
Investigation complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESTIGE ATRA GRASPER DBL-ACT 5MM
Type of Device
HANDHELD PRODUCTS & LIGATION
Manufacturer (Section D)
AESCULAP INC.
3773 corporate parkway
center valley, pa 18034
GM  18034
Manufacturer (Section G)
AESCULAP INC.
3773 corporate parkway
center valley PA 18034
Manufacturer Contact
jonathan severino
3773 corporate parkway
center valley, PA 18034
4847197287
MDR Report Key16873007
MDR Text Key314933395
Report Number2916714-2023-00037
Device Sequence Number1
Product Code NWV
UDI-Device Identifier04046955083374
UDI-Public4046955083374
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8360-10
Device Catalogue Number8360-10
Device Lot NumberM46350
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/17/2023
Initial Date FDA Received05/04/2023
Supplement Dates Manufacturer Received04/17/2023
Supplement Dates FDA Received08/23/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-