SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C1034J |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Udi section of d4 is unknown, no product information has been provided to date.D4: expiration date and h4: manufacture date are unknown, no information is available based on reported lot number.B3: month and year of event have been provided, day is unknown.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that during the pre-use check, leakage of air from the product was observed.No patient involvement.
|
|
Manufacturer Narrative
|
Other, other text: d10 - device available for evaluation, h3 - device evaluated by manufacturer, h6 - evaluation codes: updated device evaluation: one device without its original packaging was returned for investigation.Visual inspection could not find any defect that would cause leaking.Functional testing confirmed the complaint when the device failed the leak test.Root cause of this condition could have been generated during transportation and/or handling of the product.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.All mitigations currently in place were verified and it was confirmed they have been executed accordingly.They will continue to be monitored for this failure condition in this product for threshold or escalation.
|
|
Search Alerts/Recalls
|
|
|