Catalog Number UNKNOWN
Device Problem Biocompatibility (2886)
Patient Problems Burning Sensation (2146); Metal Related Pathology (4530); Implant Pain (4561)
Event Date 04/08/2022
Event Type  Injury  
Event Description
It was reported that, after undergoing r3-tha surgery on (b)(6) 2018, the patient experienced pain and a squeaking noise in the surgical area.Steroid shots were prescribed to ease the pain.This symptom persisted for two years, along with burning and sensitivity in her leg and hip, for which opioids were recommended.On (b)(6) 2022, the patient consulted these symptoms with a different surgeon, who indicated she had metal shavings and shattering in her leg.On (b)(6) 2022, a revision surgery was conducted in order to replace the socket.Patient's current health status is unknown.
 
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Manufacturer Narrative
H3.H6: the device was not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that there were no clinical factors found which would have contributed to the reported event/clinical reactions.The patient impact beyond the reported revision cannot be determined with the limited information provided.Device specific identifiers were not provided.Therefore, an evaluation of the manufacturing records, complaint history review, instructions for use, risk management file and prior actions review could not be performed.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include alignment, abnormal motion over time and/or irregular implant interaction.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN R3 SHELL
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16887786
MDR Text Key314756312
Report Number1020279-2023-01002
Device Sequence Number1
Product Code MRA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/13/2023
Initial Date FDA Received05/08/2023
Supplement Dates Manufacturer Received06/12/2023
Supplement Dates FDA Received06/13/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-