This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.Investigation of the available information determined that the minute ventilation (mv) sensor in this device has the potential to increase the pacing rate to the maximum sensor rate more quickly than expected.As a result, a software update was released in 2013 that reduced the minimum mv response factor settings and re-distributed the remaining mv response factors to provide a consistent change in sensor rate across the range of mv response factors.This corrective action is expected to reduce, but not eliminate the occurrence of this behavior.
|