Model Number LXMC17
Device Problems Device Appears to Trigger Rejection (1524); Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
Patient Problems Dehydration (1807); Dysphagia/ Odynophagia (1815); Inflammation (1932); Pain (1994)
Event Date 04/18/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).Lot number was received and dhr is pending review.When the review is completed, a supplemental medwatch will be sent with a summary of the evaluation.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: prior to linx placement, did the patient have an egd, ph, and manometry studies done? if yes, could you please share the results? when using the linx sizing device what technique was used to determine the size? did the patient have an autoimmune disease? is the patient currently taking steroids / immunosuppressive drugs? did the patient have any pre-existing dysphagia or other conditions (other than gerd)? how severe was the dysphagia/odynophagia before intervention? were there any intra-operative complications during implant? was there any hiatal or crural repair done at the same time as the implant? was the device found in the correct position/geometry at the time of removal? will the device be returning for analysis? if yes, to whom should a shipper kit be sent to? could you please share the test results of the metal allergy test? additional information was requested, and the following was obtained: did the explant take place may 1, 2023? if no, why not? if the explant was rescheduled what is the new explant date? answer - i did explant the device on monday.I will let you know how she is doing when i see her and follow up.After i removed it, i also dilated her to 20mm.There was a very intense inflammatory reaction around the device.More than i had seen before.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that a patient had linx lxmc17 implanted on (b)(6) 2023 and on (b)(6) 2023 patient started experiencing severe dysphagia and sharp chest pains.Md put on steroid prescription- with no relief.On (b)(6) 2023 patient admitted to er was dehydrated and in severe chest pain.Possibility of undetected metal allergy- which was inconclusive.Md conducted a dilation on patient on (b)(6) 2023.At this time patient cannot tolerate it in them and is back in er.Linx device is scheduled to be removed (b)(6) 2023.
 
Manufacturer Narrative
(b)(6).Date sent: 6/14/2023.Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.However, if the product is received at a later date, the investigation will be updated as applicable.A manufacturing record evaluation was performed for the finished device lot number: 27390, and no related nonconformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5T LINX, 17 BEADS
Type of Device
ANTI-REFLUX IMPLANT
Manufacturer (Section D)
TORAX MEDICAL, INC.
4545 creek road
cincinnati OH
Manufacturer Contact
kate karberg
4188 lexington avenue north
shoreview 
3035526892
MDR Report Key16894530
MDR Text Key314817373
Report Number3008766073-2023-00080
Device Sequence Number1
Product Code LEI
UDI-Device Identifier00855106005370
UDI-Public00855106005370
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/27/2024
Device Model NumberLXMC17
Device Catalogue NumberLXMC17
Device Lot Number27390
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/28/2023
Initial Date FDA Received05/09/2023
Supplement Dates Manufacturer Received06/14/2023
Supplement Dates FDA Received06/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/27/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-