MEDTRONIC NAVIGATION, INC STEALTHSTATION FLEXENT; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT
|
Back to Search Results |
|
Model Number 9736242 |
Device Problem
Imprecision (1307)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/03/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H3, h6) system checkout has not begun, nor have any parts been returned to the manufacturer for analysis.Codes b17, c20, and d15 are applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation system used intraoperatively for a functional endoscopic sinus surgery (fess) in which there was there was patient involvement.The doctor reported they had a very good trace, yellow circle (> 2 millimeter accuracy) around the whole head, and felt it was accurate once verifying registration.The surgeon got a 2.4 millimeter on the registration accuracy.However, when they were touching the back wall of the sinus, the surgeon indicated the navigation showed they had not hit the wall yet.Navigation was 1.5 centimeters off consistently anteriorly.Medtronic imaging was aborted.The surgery continued, there was no impact to patient, and surgical delay time was less than one-hour.
|
|
Search Alerts/Recalls
|
|
|