Model Number 867030
Device Problem Low Readings (2460)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/12/2023
Event Type  malfunction  
Event Description
The customer reported the spo2 value displayed on the x3 (b06sp6) for bed 9 is too low.The device was in use at time of event, there was no adverse event reported.
 
Manufacturer Narrative
A philips field service engineer (fse) went onsite to evaluate the devices in question.The testing conducted, the cause of the reported problem was confirmed.The reported problem was confirmed to be the spo2 parameter board.After spo2 parameter board replacement, the device was returned to functional use with no further issues identified.The device remains at the customer site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE X3
Type of Device
INTELLIVUE X3
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key16898721
MDR Text Key314885992
Report Number9610816-2023-00219
Device Sequence Number1
Product Code MHX
UDI-Device Identifier00884838082588
UDI-Public00884838082588
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K171801
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number867030
Device Catalogue Number867030
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/12/2023
Initial Date FDA Received05/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/17/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-