Catalog Number UNK_WWA
Device Problems Inadequacy of Device Shape and/or Size (1583); Adverse Event Without Identified Device or Use Problem (2993); Osseointegration Problem (3003)
Patient Problems Loss of Range of Motion (2032); Inadequate Osseointegration (2646); Implant Pain (4561)
Event Date 04/12/2023
Event Type  Injury  
Manufacturer Narrative
The device is not available for evaluation as it remains implanted in the patient.A review of the device history is not possible because the lot number was not communicated.If additional information becomes available, it will be provided on a supplemental report.H3 other text : device remains implanted in patient.
 
Event Description
It was reported that the patient underwent a total ankle replacement.Allegedly, the patient may need to undergo a revision surgery of both tibia and the talus.Clinical reason is pain, stiffness of ankle, loosening of tibia and impingement of the talar component at gutters.
 
Manufacturer Narrative
Correction - h6 (device code, results code and conclusion code).The reported event could be confirmed, since evaluation of ct imaging finds the implant is too large for the joint space.Upon further investigation of the ct scans by healthcare professionals the following was observed, ¿the talar component shows some subsidence at the anterolateral aspect.That can only happen if there is an osseointegration problem (loosening) and migration.The talar component is described as being too large.The ct scans would support this opinion." based on investigation, the root cause was attributed to a combination of user and patient related issue.The event was caused by implanting surgeon using a talar component too large and also osseointegration problems indicative of poor bone quality.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the lot number was not communicated.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event Description
It was reported that the patient underwent a total ankle replacement.Allegedly, the patient may need to undergo a revision surgery of both tibia and the talus.Clinical reason is pain, stiffness of ankle, loosening of tibia and impingement of the talar component at gutters.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN TALAR COMPONENT
Type of Device
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16899968
MDR Text Key314884332
Report Number3010667733-2023-00245
Device Sequence Number1
Product Code HSN
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Catalogue NumberUNK_WWA
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/12/2023
Initial Date FDA Received05/09/2023
Supplement Dates Manufacturer Received08/03/2023
Supplement Dates FDA Received08/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-