MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. OES CYSTONEPHROFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. OES CYSTONEPHROFIBERSCOPE Back to Search Results
Model Number CYF-5
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/11/2023
Event Type  malfunction  
Event Description
The customer reported the oes cystonephrofiberscope rubber on the flexible end near the tip was ripped.The event occurred during reprocessing and there was no patient harm or user injury reported due to the event.
 
Manufacturer Narrative
The device was returned for investigation.Upon evaluation of the device, bending section cover crack, bending section cover leakage, control body insulation, control body burned, and image centering off were observed.The investigation is ongoing, therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The customer's complaint was confirmed.Since the device is more than 15 years old, the device history record was unable to be reviewed.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, it is likely that the reportable phenomenon occurred due to the handling methods of the facility.The event can be prevented by following the instructions for use which state: "·perform a leakage test on the endoscope after each precleaning procedure.Do not use the endoscope if a leak is detected.Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions.Use of a leaking endoscope may also pose an infection control risk.·the accessories listed on the front cover of this manual cannot be refurbished or repaired and are intended to be replaced once they show any signs of wear and tear.Should any irregularity be observed, use a replacement accessory instead.Using defective accessories may cause equipment malfunction, reduce the efficacy of reprocessing, present a risk to patients and/or operators, or damage the endoscope and/or accessories.·single-use brush, such as the single use combination cleaning brush (bw-411b), is designed for cleaning only one endoscope and its related accessories.Dispose of the single-use brush immediately after use.Using a single-use brush to clean multiple endoscopes and/or accessories may reduce its cleaning efficacy and may damage the brush leading to brush breakage or the endoscope and/or accessory damage." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CYSTONEPHROFIBERSCOPE
Type of Device
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16904223
MDR Text Key315365330
Report Number3002808148-2023-04713
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170339417
UDI-Public04953170339417
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K221690
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 06/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/11/2023
Initial Date FDA Received05/10/2023
Supplement Dates Manufacturer Received06/03/2023
Supplement Dates FDA Received06/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/17/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-