Model Number CYF-5 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/11/2023 |
Event Type
malfunction
|
Event Description
|
The customer reported the oes cystonephrofiberscope rubber on the flexible end near the tip was ripped.The event occurred during reprocessing and there was no patient harm or user injury reported due to the event.
|
|
Manufacturer Narrative
|
The device was returned for investigation.Upon evaluation of the device, bending section cover crack, bending section cover leakage, control body insulation, control body burned, and image centering off were observed.The investigation is ongoing, therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The customer's complaint was confirmed.Since the device is more than 15 years old, the device history record was unable to be reviewed.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, it is likely that the reportable phenomenon occurred due to the handling methods of the facility.The event can be prevented by following the instructions for use which state: "·perform a leakage test on the endoscope after each precleaning procedure.Do not use the endoscope if a leak is detected.Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions.Use of a leaking endoscope may also pose an infection control risk.·the accessories listed on the front cover of this manual cannot be refurbished or repaired and are intended to be replaced once they show any signs of wear and tear.Should any irregularity be observed, use a replacement accessory instead.Using defective accessories may cause equipment malfunction, reduce the efficacy of reprocessing, present a risk to patients and/or operators, or damage the endoscope and/or accessories.·single-use brush, such as the single use combination cleaning brush (bw-411b), is designed for cleaning only one endoscope and its related accessories.Dispose of the single-use brush immediately after use.Using a single-use brush to clean multiple endoscopes and/or accessories may reduce its cleaning efficacy and may damage the brush leading to brush breakage or the endoscope and/or accessory damage." olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|