Model Number 384539
Device Problem Crack (1135)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
¿our nicu uses cath picc 1lum 1.9fr manufacturing #384539 and we have had repeated concerns with cracking at the hub and leaking.This issue has led to several lines needing to be replaced or repaired.Have you had other facilities bring up this concern?¿.
 
Manufacturer Narrative
According to the product experience report, there was no sample to be returned.Additionally, no photographic or visual evidence of any kind was provided which would have allowed for the allegation to be reviewed.Several complaints for this part number have previously been received regarding a cracked hub resulting in leakage, and capa 2021-039 was initiated to address this issue.The capa is currently in the implementation phase and will evaluate the corrective action implementation for effectiveness to prevent a recurrence of this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
L-CATH PICC S/L 26GA (1.9F) 0.60MM X 30CM
Type of Device
L-CATH PICC
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX 75751
Manufacturer Contact
elnaz rahman
1445 flat creek rd
athens, TX 75751
9036759321
MDR Report Key16904466
MDR Text Key314932256
Report Number0001625425-2023-01052
Device Sequence Number1
Product Code LJS
UDI-Device Identifier00886333209934
UDI-Public00886333209934
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091670
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 05/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number384539
Device Catalogue Number384539
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/26/2023
Initial Date FDA Received05/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-