• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 X-PORT ISP IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 X-PORT ISP IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 7707540J
Device Problems Fracture (1260); Material Separation (1562); Dent in Material (2526)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/13/2023
Event Type  malfunction  
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the x-port isp implantable port, groshong single-lumen, 8f products that are cleared in the us.The pro code and 510 k number for the x-port isp implantable port, groshong single-lumen, 8f products are identified in d2 and g4.H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that some time post a port placement, the catheter was allegedly broken.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the x-port isp implantable port, groshong single-lumen, 8f products that are cleared in the us.The pro code and 510 k number for the x-port isp implantable port, groshong single-lumen, 8f products are identified in d2 and g4.H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one x-port implantable port attached to a groshong catheter was received for evaluation.Visual, microscopic, functional and tactile evaluations were performed.A complete circumferential break was noted on the proximal end of the distal catheter segment and the distal end of the attached catheter which appeared elliptical in shape.The edges of the breaks were noted to be uneven.Small splits were observed on the border of both break surfaces and were noted to be granular.Therefore, the investigation is confirmed for the reported fracture and the identified material separation and dent issues.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event Description
It was reported that some time post a port placement, the catheter was allegedly broken.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-PORT ISP IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key16909320
MDR Text Key314977571
Report Number3006260740-2023-01808
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741098246
UDI-Public(01)00801741098246
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K022983
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number7707540J
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/14/2023
Initial Date FDA Received05/11/2023
Supplement Dates Manufacturer Received07/12/2023
Supplement Dates FDA Received07/14/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-