C.R. BARD, INC. (BASD) -3006260740 X-PORT ISP IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 7707540J |
Device Problems
Fracture (1260); Material Separation (1562); Dent in Material (2526)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/13/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the x-port isp implantable port, groshong single-lumen, 8f products that are cleared in the us.The pro code and 510 k number for the x-port isp implantable port, groshong single-lumen, 8f products are identified in d2 and g4.H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that some time post a port placement, the catheter was allegedly broken.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the x-port isp implantable port, groshong single-lumen, 8f products that are cleared in the us.The pro code and 510 k number for the x-port isp implantable port, groshong single-lumen, 8f products are identified in d2 and g4.H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one x-port implantable port attached to a groshong catheter was received for evaluation.Visual, microscopic, functional and tactile evaluations were performed.A complete circumferential break was noted on the proximal end of the distal catheter segment and the distal end of the attached catheter which appeared elliptical in shape.The edges of the breaks were noted to be uneven.Small splits were observed on the border of both break surfaces and were noted to be granular.Therefore, the investigation is confirmed for the reported fracture and the identified material separation and dent issues.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|
|
Event Description
|
It was reported that some time post a port placement, the catheter was allegedly broken.There was no reported patient injury.
|
|