ZIMMER BIOMET, INC. G7 DEPTH GAUGE; HIPS, INSTRUMENTS Back to Search Results
Model Number N/A
Device Problems Fracture (1260); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/12/2023
Event Type  malfunction  
Event Description
It was reported the piece came in extremely bent.An attempt to bend it back in place was made and it broke immediately.Attempts have been made and no further information has been provided.No consequences or impact to the patient.
 
Manufacturer Narrative
(b)(4).Visual examination of the returned product identified the device had fractured at one of the measurement notches.There is scuffing to the body of the device.Complaint confirmed based on evaluation of returned product.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 DEPTH GAUGE
Type of Device
HIPS, INSTRUMENTS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16910961
MDR Text Key314995475
Report Number0001825034-2023-01022
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 05/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110010717
Device Lot Number421222
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/19/2023
Initial Date FDA Received05/11/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/09/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-