SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX EXPANDABLE BREATHING CIRCUITS, PEDIATRIC; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C49101320J |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during the pre-use check, leakage of air from the breathing circuit was observed.No adverse patient effects were reported by the customer.It has been reported that no additional information is available for this event.
|
|
Manufacturer Narrative
|
Month and year of event have been provided, day is unknown.D4: udi section is blank, it is unavailable for this catalog number.G5: 510k is blank, device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
|
|
Manufacturer Narrative
|
One device and two photos were returned for investigation.The returned sample was visually inspected and did not confirm any air leak defect.Then, the leak test was performed.The leaking defect was not detected in the leak test; complaint was not confirmed.No root cause could be determined since the complaint was not confirmed due sample provided does not shows failure mode reported.No corrective actions are required since the complaint was not confirmed.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
|
|