AESCULAP AG CASPAR RONGEURSTRSERR 3X12MM160MM; BONE PUNCHES, RONGEURS Back to Search Results
Model Number FF533R
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/20/2023
Event Type  Injury  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event Description
It was reported that there was an issue with ff533r - caspar rongeurstrserr 3x12mm160mm.According to the complaint description, the discectomy forceps break in the operating area during a spinal surgery.The broken jaw was removed.The surgeon requested a second discectomy forceps to continue the procedure.The second forceps also broke.The jaw broke a little further away in the operating area and it was removed.An additional medical intervention was necessary.Additional information was not provided nor available.Additional patient information is not available.The adverse event is filed under aag reference (b)(4).
 
Manufacturer Narrative
Additional information: d9 - product return.H3 - yes, evaluation.H6 - codes updated.Investigation results: aesculap ag carried out a visual and microscopic examination.A repair stamp was found, indicating that the instrument had been repaired in france by aesculap technical services (ats).The movable jaw was broken off - the fragment, which could provide further clues as to the reason for the deviation, is not available to us.The fracture surface shows a forced fracture due to overloading.The hardness was tested and found to be 471 hardness according to vickers (hv) 5, which is within the specification of 420 +110 hv5.Batch history review: the lot number was not provided.A review was not performed.According to manufacturer's reporting evaluation in accordance with 21 cfr part 803, section 803.3, this event is considered reportable for the following reason - adverse event (not later than 30 days).The assessment for the reportability of this adverse event was based on the patient harm, additional medical intervention.Conclusion/preventive measures: based upon the investigation results, a definitive root cause cannot be determined.The results show no hints for a product or manufacturing error.Notes from the electronic instructions for use eifu) ta011658 2020-10 v6 change no.62252 on this: damage or destruction of the jaws due to overload! only use rongeurs as intended (i.E.Ablation of soft tissue).Do not cut or lever out bones or adjacent bone structures.Avoid overloading the jaws by tearing off or holding and levering tissue or by lateral loading through twisting, lateral strain due to twisting.Do not use rongeurs with blunt cutting edges, send them in for repair.Do not remove more fabric than will fit into the openings of the jaws.Based upon the investigation results, a capa is not required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CASPAR RONGEURSTRSERR 3X12MM160MM
Type of Device
BONE PUNCHES, RONGEURS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key16912681
MDR Text Key315019093
Report Number9610612-2023-00111
Device Sequence Number1
Product Code HTX
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFF533R
Device Catalogue NumberFF533R
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/02/2023
Initial Date FDA Received05/11/2023
Supplement Dates Manufacturer Received04/29/2024
Supplement Dates FDA Received05/23/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-