• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PERFUSION SYSTEMS BIO-CONSOLE; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PERFUSION SYSTEMS BIO-CONSOLE; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 560
Device Problems Mechanical Problem (1384); Material Deformation (2976); Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/10/2021
Event Type  malfunction  
Manufacturer Narrative
Device evaluation summary: during preventive maintenance by a service technician it was noted this bio-console instrument displayed an error message (power failure) although the instrument was always connected to the mains.The service technician noted that when the motor speed was increased a message came up with ac power failure and they observed that the output voltage of the power supply was out of specification and it could not be adjusted (26.7v and drops down with a motor speed into 22.8v).A pressure leakage was observed in channel 1 and 2 and they observed a broken blind connector inside the base unit of the instrument.The issues were resolved by replacing the power supply ), blind connector, motor / pressure board and hw screw ph ss pf 1/2 and performing a technical services updated to update the user interface (ui) software to version 2.U02.010.Preventive maintenance was completed per specifications.Conclusion: complaint confirmed.The service history record was not reviewed as returned product analysis found no evidence of servicing issues with the serviced device.There were no patient/clinical safety issues reported.Trends for issues with this product are reviewed at quarterly quality meetings.This regulatory report is being submitted as part of a retrospective review and remediation per d00953163 as part of a capa action.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
During preventive maintenance by a service technician it was noted this bio-console instrument displayed an error message (power failure) although the instrument was always connected to the mains.The service technician noted that when the motor speed was increased a message came up with ac power failure and they observed that the output voltage of the power supply was out of specification and it could not be adjusted (26.7v and drops down with a motor speed into 22.8v).A pressure leakage was observed in channel 1 and 2 and they observed a broken blind connector inside the base unit of the instrument.This was detected during service so there was no adverse patient effect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-CONSOLE
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16918856
MDR Text Key315083685
Report Number2184009-2023-00257
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K070286
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number560
Device Catalogue Number560
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/10/2021
Initial Date FDA Received05/12/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-