E1: initial reporter facility name: (b)(6) hospital.Device evaluated by mfr: returned product consisted of an solent omni thrombectomy.The pump, effluent/supply line, shaft, tip, piston, and spike line were visually inspected.Visual examination revealed that multiple kinks along the shaft.Microscopic examination revealed that the hypotube is separated 25.5cm from the tip.Functional testing was performed by placing the device in the angiojet ultra console.The device did not prime and there was a check saline supply error.Inspection of the device presented no other damage or irregularities.Product analysis confirmed the hypotube is separated which would cause the device to have and error and leak.
|
Reportable based on device analysis completed on (b)(6) 2023.It was reported that pump seal fault occurred.The target location was located in the deep vein thrombosis of the left lower extremity.An angiojet solent omni was used for a thrombectomy procedure.During the procedure while in thrombectomy mode for 30 seconds, it was found that there was liquid in the air sac of the pump resulting to insufficient suction and affecting the procedure.The procedure was completed with another of the same device.No patient complications were reported, and the patient was stable.However, returned device analysis revealed a hypotube break.
|