Catalog Number 08828601190
Device Problem Incorrect Measurement (1383)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/10/2023
Event Type  malfunction  
Manufacturer Narrative
Common device name continued: cancer from benign conditions calibrations performed on (b)(6) 2023 had signals below the expected range.Quality controls were acceptable.Instrument performance testing was acceptable.The investigation is ongoing.
 
Event Description
The initial reporter stated they received questionable results for one patient tested with the elecsys free psa and the elecsys total psa assay on a cobas e411 analyzer.The free psa values were greater than the total psa values.The questionable results were not reported outside of the laboratory.This medwatch will apply to the free psa assay.Please refer to the medwatch with patient identifier (b)(6) for information related to the total psa assay.A first sample of the patient was tested on (b)(6) 2023: the first sample initially resulted in a free psa value of 1.46 ng/ml.The first sample initially resulted in a total psa value of 0.066 ng/ml.A second sample of the patient was tested on (b)(6) 2023: the second sample initially resulted in a free psa value of 1.58 ng/ml.The second sample initially resulted in a total psa value of 0.061 ng/ml.A third sample of the patient was tested on (b)(6) 2023 in a different laboratory using the siemens method: the third sample initially resulted in a free psa value of 0.03 ng/ml.The third sample initially resulted in a total psa value of 1.21 ng/ml.The results obtained with the siemens method are considered correct.The serial number of the e411 analyzer is (b)(6).
 
Manufacturer Narrative
The patient sample was requested for investigation but could not be provided.The investigation could not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FREE PSA
Type of Device
TEST, PROSTATE SPECIFIC ANTIGEN, FREE, (NONCOMPLEXED) TO DISTINGUISH PROSTATE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16919287
MDR Text Key315098274
Report Number1823260-2023-01615
Device Sequence Number1
Product Code MTG
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
P000027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2023
Device Catalogue Number08828601190
Device Lot Number617615
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/25/2023
Initial Date FDA Received05/12/2023
Supplement Dates Manufacturer Received06/23/2023
Supplement Dates FDA Received07/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age54 YR
Patient SexMale
-
-