The patient was undergoing a thrombectomy procedure in the right m2 segment of the middle cerebral artery (mca) using a penumbra system 4max reperfusion catheter (4maxc), a non-penumbra sheath and a guidewire.It should be noted that the patients anatomy was tortuous.During the procedure, the physician successfully completed two passes with the 4maxc.Next, while advancing the 4maxc over the guidewire towards the target location, the physician experienced resistance and decided to remove the 4maxc.Upon removal of the 4maxc, the physician found that the 4maxc was fractured at the mid-shaft and connected by the inner wire braiding.Therefore, the 4maxc was no longer used in the procedure.The procedure was completed using a new 4maxc, a new guidewire, and the same sheath.There was no report of an adverse effect to the patient.
|
Evaluation of the returned 4maxc confirmed that the catheter was fractured.If the 4maxc is advanced against resistance, damage such a kink and subsequent fracture may occur.Based on the reported complaint, the patient anatomy was tortuous.The tortuous anatomy may have contributed to resistance during advancement.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|