• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPORT CLEARVUE ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPORT CLEARVUE ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 5608062
Device Problem Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/17/2023
Event Type  malfunction  
Event Description
It was reported that during preparation of a port placement procedure, the port allegedly had cuts and sterility of the device was compromised.There was no patient contact.
 
Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records will be performed.The device has been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.H10: d4 (expiry date: 11/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that upon receiving the device, the port package allegedly had cuts and sterility of the device was compromised.There was no patient contact.
 
Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one sealed powerport clearvue isp implantable port kit was received for evaluation and two photos were provided for review.Visual evaluation was performed.A longitudinal split was noted on the top center portion of the outer packaging and the split appeared to go within the product tray label.Manufacturing site evaluation of the sample found a longitudinal cut at the bottom of the unit label placement, the back of the tray and the inner packaging appeared to be free of damage, the condition of the shipping box is unknown.Therefore, the investigation is confirmed for the reported packaging of the device torn issue and the photo review also confirms the same.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 11/2024), g3, h6 (method).The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT CLEARVUE ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key16928546
MDR Text Key315471438
Report Number3006260740-2023-01874
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741027369
UDI-Public(01)00801741027369
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5608062
Device Catalogue Number5608062
Device Lot NumberREHP1304
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/17/2023
Initial Date FDA Received05/15/2023
Supplement Dates Manufacturer Received10/23/2023
Supplement Dates FDA Received10/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-