Catalog Number DL950F
Device Problems Break (1069); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/24/2022
Event Type  malfunction  
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the denali products that are cleared in the us.The pro code and 510 k number for the denali products are identified in d2 and g4.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.H10: d4 (expiration date: 06/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during a vena cava filter placement procedure, when the filter was advanced during implantation, the introducer sheath allegedly broke.The procedure was completed by using another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the denali products that are cleared in the us.The pro code and 510 k number for the denali products are identified in d2 and g4.H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.Two electronic photos were provided and reviewed.The first photo & second photos shows the introducer sheath, the dilator, the pusher catheter, and the filter outside of the delivery system.Blood residue was noted.The introducer sheath was noted to be completely break and detached into two piece in the package.Therefore, the investigation is confirmed for the reported detachment issue.A definitive root cause for the reported detachment issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiration date: 06/2025), g3, h6 (method) h11: h6 (device, result, conclusion) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during a vena cava filter placement procedure via femoral access, when the filter was advanced during implantation, the introducer sheath allegedly broke.The procedure was completed by using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DENALI FEMORAL SYSTEM
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key16929033
MDR Text Key315190329
Report Number2020394-2023-00332
Device Sequence Number1
Product Code DTK
UDI-Device Identifier00801741040825
UDI-Public(01)00801741040825
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K130366
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberDL950F
Device Lot NumberGFGT0350
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/17/2023
Initial Date FDA Received05/15/2023
Supplement Dates Manufacturer Received06/22/2023
Supplement Dates FDA Received06/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-