• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EASYTRAK 2; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EASYTRAK 2; IMPLANTABLE LEAD Back to Search Results
Model Number 4518
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/26/2023
Event Type  Injury  
Event Description
It was reported that this left ventricular (lv) lead was explanted due to it was fractured.It was confirmed through a fluoroscopy.A new left ventricular (lv) lead was successfully implanted.No additional adverse patient effects were reported.
 
Event Description
It was reported that this left ventricular (lv) lead was explanted due to it was fractured.It was confirmed through a fluoroscopy.A new left ventricular (lv) lead was successfully implanted.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Testing was completed to assess lead electrical performance.Measurements throughout this test were within normal limits.Laboratory testing was unable to reproduce the reported clinical observations and detailed analysis did not reveal any abnormalities.Laboratory analysis did not identify any lead characteristics that would have caused or contributed to the reported clinical observations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16938706
MDR Text Key315292679
Report Number2124215-2023-24241
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526417672
UDI-Public00802526417672
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P010012/S024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/18/2011
Device Model Number4518
Device Catalogue Number4518
Device Lot Number336880
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/26/2023
Initial Date FDA Received05/16/2023
Supplement Dates Manufacturer Received07/20/2023
Supplement Dates FDA Received07/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/18/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age65 YR
Patient SexFemale
-
-