• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES Back to Search Results
Model Number CDS0706-XTW
Device Problems Break (1069); Unstable (1667); Difficult to Open or Close (2921); Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/26/2023
Event Type  Injury  
Event Description
This is filed to report a clip opening while locked (cowl) and clip jumping, requiring intervention.It was reported that a mitraclip procedure was performed to treat degenerative mitral regurgitation (mr) grade 4.After lock line removal during the deployment process, the clip jumped open from fully closed to approximately 40 degrees.Troubleshooting was performed but was unsuccessful.Subsequently, the physician deployed the clip.It was noted that the arm positioner rotated freely, and the clip no longer responded to communication; a cable break was suspected.A second clip was then implanted lateral to the cowl clip for stabilization.The procedure was concluded with a resulting mr of grade <1.There was no clinically significant delay in the procedure and no adverse patient sequelae.No additional information was provided.
 
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
N/a.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified did not indicate a lot-specific quality issue.Based on available information and without the device to analyze, the cause of the reported clip opening while locked & clip arms jumping during the procedure could not be determined.The cause of the reported unstable (arm positioner) & break (cable break) cannot be determined.The reported unexpected medical intervention was a result of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key16938958
MDR Text Key315298310
Report Number2135147-2023-02123
Device Sequence Number1
Product Code NKM
UDI-Device Identifier05415067037381
UDI-Public05415067037381
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/22/2024
Device Model NumberCDS0706-XTW
Device Catalogue NumberCDS0706-XTW
Device Lot Number30119R1098
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/26/2023
Initial Date FDA Received05/16/2023
Supplement Dates Manufacturer Received06/13/2023
Supplement Dates FDA Received06/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/23/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER.
Patient Outcome(s) Required Intervention;
-
-