MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC STEALTHSTATION¿ S8 SYSTEM; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC STEALTHSTATION¿ S8 SYSTEM; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735669
Device Problems Imprecision (1307); Human-Device Interface Problem (2949)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/27/2023
Event Type  malfunction  
Event Description
Medtronic received information regarding a navigation system.It was reported that during surgery, the surgeon called to report that they were having a hard time passing registration.When navigating, the surgeon claimed that the tracking was consistently 0.5cm off.Surgery continued.It was noted that while registering, surgeon did not have the best technique, 3d model also did not have the best quality or the top of the head.Site claimed that a lot of surgeon struggles on registration on this specific system compared to other system on site.
 
Manufacturer Narrative
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 9 735736, serial/lot #: (b)(6).H3: no parts have been received by the manufacturer for evaluation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H3, h6: a medtronic representative went to the site to test the equipment.Testing revealed that the system passed testing.There was no failures found.Codes b01, c19 and d14 are applicable to this analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H3,h6: a software analysis was initiated.However, the software evaluation found that a probable cause was unable to be determined due to insufficient information.Evaluation codes that apply to this testing: b01, c19, d15 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information in sections a and b5.Correction to d2.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received.It was reported that this occurred during a functional endoscopic sinus surgery (fess).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION¿ S8 SYSTEM
Type of Device
EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key16939823
MDR Text Key315367749
Report Number1723170-2023-00792
Device Sequence Number1
Product Code PGW
UDI-Device Identifier00643169838918
UDI-Public00643169838918
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 08/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735669
Device Catalogue Number9735669
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/27/2023
Initial Date FDA Received05/16/2023
Supplement Dates Manufacturer Received05/23/2023
07/12/2023
08/08/2023
Supplement Dates FDA Received06/01/2023
07/19/2023
08/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/30/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age49 YR
Patient SexMale
Patient Weight68 KG
-
-