• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE G PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE G PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-52
Device Problem Component Incompatible (1108)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/05/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported during a routine check getinge fse the cardiosave intra-aortic balloon pump (iabp) drive manifold needs replacing.There was no patient involvement reported.
 
Manufacturer Narrative
Updated fields: e1(site state), h6 (type of investigation, investigation findings, component codes, investigation conclusion).Additional information: contact person e-mail address: (b)(6).Event site postal code:(b)(6).Event site state: (b)(6).The getinge field service engineer (fse) that encountered the reported issue during the preventative maintenance (pm) replaced the drive manifold to fix the issue and performed a full functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released and cleared for clinical service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE G PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16940948
MDR Text Key315425340
Report Number2249723-2023-02373
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108438
UDI-Public10607567108438
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-52
Device Catalogue Number0998-00-0800-52
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 05/05/2023
Initial Date FDA Received05/16/2023
Supplement Dates Manufacturer Received02/14/2024
Supplement Dates FDA Received02/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-