Catalog Number 114362
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/19/2023
Event Type  malfunction  
Manufacturer Narrative
Initial reporter first name: (b)(6).Initial reporter last name: (b)(6).Initial reporter facility name: (b)(6).Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported an external fluid leak was observed originating from an unspecified location of an ak 96 machine during setup.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The actual device was not available; however, the device was evaluated on site by a qualified baxter technician.During visual inspection, the flush valve (flva), chemical bypass valve (cbva) and filter valve (fiva) were observed to be aging.The reported condition was verified.The cause of the condition was due to aging parts.To rectify the issue the technician replaced the parts.Post replacement, self-tests were successfully performed.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AK 96 MACHINES
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - LUND MONITORS
magistratsvagen 16
lund skane lan SE-22 643
SW   SE-22643
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16943469
MDR Text Key316020922
Report Number9616026-2023-00033
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number114362
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/19/2023
Initial Date FDA Received05/16/2023
Supplement Dates Manufacturer Received06/20/2023
Supplement Dates FDA Received07/17/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-