STRYKER ORTHOPAEDICS-MAHWAH SIZE 7 ACCOLADE II 127 DEG; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6721-0737
Device Problem Loss of Osseointegration (2408)
Patient Problem Inadequate Osseointegration (2646)
Event Date 04/24/2023
Event Type  Injury  
Event Description
As reported: "hip revision.Revised stryker implants: explanted accolade stem, femoral head, & x3 insert due to loosening of femoral stem.".
 
Manufacturer Narrative
Reported event an event regarding loosening involving an accolade stem was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to stem loosening.The event was not confirmed and the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIZE 7 ACCOLADE II 127 DEG
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key16944620
MDR Text Key315354420
Report Number0002249697-2023-00543
Device Sequence Number1
Product Code LPH
UDI-Device Identifier04546540669520
UDI-Public04546540669520
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K143085
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 05/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number6721-0737
Device Catalogue Number6721-0737
Device Lot Number85407301
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/24/2023
Initial Date FDA Received05/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age59 YR
Patient SexFemale
-
-