The device was returned for investigation.Upon evaluation of the device, fading olympus logo, peeling cements, bending section cover chip/crack, and scope connector missing lock pin at the valve were observed.The investigation is ongoing, therefore, the root cause of the reported event cannot be determined at this time,.However, if additional information becomes available, this report will be supplemented accordingly.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the cause of the event is likely due to physical stress.The event can be prevented by following the instructions for use (ifu) which state: warning do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.Olympus will continue to monitor field performance for this device.
|