Model Number CPHV
Device Problem Insufficient Information (3190)
Patient Problem Low Blood Pressure/ Hypotension (1914)
Event Date 04/18/2023
Event Type  Death  
Manufacturer Narrative
A complete manufacturing and material records review for the carbomedics top hat mechanical valve, model #s5-021 s/n #(b)(6) 2023, has been performed.The results confirmed that the device satisfied all material, visual and performance standards required at the time of manufacture and release.Since minimal information was shared with the manufacturer and the device was not returned for analysis, a definitive root cause of the reported event cannot be established.However, from the document review performed, no manufacturing deficiencies were identified.The manufacturer has been informed that no further information will be shared by the site.The manufacturer will provide a follow up report in case any additional information is received.H3 other text : unknown disposition.
 
Event Description
The manufacturer was informed that on (b)(6) 2023 a 51 year old patient underwent avr with carbomedics top hat supra-annular aortic prosthesis size 21.The surgical procedure was successfully completed.On first post-operative day, the patient experienced a sudden decrease in blood pressure and heart rate couldn't be measured.Ultimately, the patient passed away.No information is available to the manufacturer regarding patient clinical history and risk factors, details on surgical procedure (e.G.If concomitant procedures were performed).No data are available to the manufacturer regarding device functionality post-op.It was reported that the surgeon suspected that patient death was caused by leaflet blockage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARBOMEDICS TOP HAT MECHANICAL HEART VALVE
Type of Device
MECHANICAL HEART VALVE PROSTHESIS
Manufacturer (Section D)
CORCYM S.R.L.
strada crescentino
saluggia, vercelli 13040
IT  13040
Manufacturer (Section G)
CORCYM S.R.L.
strada crescentino
saluggia, vercelli 13040
IT   13040
Manufacturer Contact
laura mannino
5005 north fraser way
burnaby, bc V5J 5-M1
CA   V5J 5M1
MDR Report Key16946724
MDR Text Key315389384
Report Number3005687633-2023-00113
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier08022057012906
UDI-Public(01)08022057012906(240)S5-021(17)270419
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P900060
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberCPHV
Device Catalogue NumberS5-021
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/18/2023
Initial Date FDA Received05/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age51 YR
Patient SexMale
Patient Weight74 KG
-
-